Vital Signs Monitor User Manual Product Specification
- 138 -
Adult Mode
120 mmHg, 140 mmHg, 150 mmHg, 160 mmHg, 180
mmHg, 200 mmHg, 220 mmHg, 240 mmHg, 260 mmHg,
280 mmHg
Default: 160 mmHg
Pediatric Mode
80 mmHg, 100 mmHg, 120 mmHg, 140 mmHg, 150
mmHg, 160 mmHg, 180 mmHg, 200 mmHg, 220 mmHg,
250 mmHg
Default: 120 mmHg
Neonatal Mode
60 mmHg, 70 mmHg, 80 mmHg, 90 mmHg, 100 mmHg,
120 mmHg, 140 mmHg
Default: 90 mmHg
A.4 SpO
2
Complies with ISO 80601-2-61: 2017.
EDAN Module
Measuring Range
0% to 100%
Resolution 1%
☆
Data Update Period
1 s
☆
Accuracy
☆
Adult /Pediatric
±
2% (70% to 100% SpO
2
)
Undefined (0% to 69% SpO
2
)
☆
Neonate
±
3% (70% to 100% SpO
2
)
Undefined (0% to 69% SpO
2
)
SpO
2
Storage Interval
In Ward round or spot-checking mode
30 S (default), 1 min, 2 mins, 5 mins
Sensor
Red Light
(660
±
3) nm
Infrared Light
(905
±
10) nm
Emitted Light Energy
< 15 mW
PI
Measuring Range
0-10, invalid PI value is 0.
Resolution 1