Vital Signs Monitor User Manual Alarm Information
- 64 -
Message
Cause
Alarm Level
Action Taken
Network Disconnect
In distributed alarm
system, the monitor’s
network is disconnected.
Low
1) Check if the
network cable is
well connected.
2) Check if the
MFM-CMS is
turned on.
3) Check if the IP
of bedside monitor
and MFM-CMS are
on the same
network segment.
4) Contact the
professionals
authorized by
manufacturer to
check the network
problem.
Network Traffic
Abnormity
Abnormal network traffic
has been detected. The
data traffic exceeds the
limit.
High
Disconnect the
network to make
the monitor work
properly, and then
contact the
professionals
authorized by
manufacturer to
check the network
problem.
9.3 Prompts
Message
Cause
SpO
2
Search Pulse
SpO
2
module is analyzing the patient signal and searching
for the pulse to compute the saturation, when sensor is
connected with patient.
SpO
2
Noisy Signal
There is interference with SpO
2
measurement signals due
to patient movement, ambient light, electrical interference
or else. (Nellcor SpO
2
)
Manual Measuring
In manual measuring mode
Continual Measuring
In continuous measuring mode