Vital Signs Monitor User Manual Monitoring TEMP
- 90 -
If the predict measuring is successfully finished, the monitor enters
MONITOR
mode after 30s;
otherwise the monitor enters
MONITOR
mode immediately after the predict measuring. The
monitoring state lasts for 10 min, and then the monitor enters waiting state.
— — —
displays in the
TEMP parameter area on interface. Put the sensor back into the sensor bracket.
If necessary, repeat the measurement according to the procedure above.
NOTE:
1. After one measurement, the user should put the sensor back to the sensor bracket
and then take it out for starting a new measurement.
2. The reference body site temperature is the same as the temperature of the measuring
site.
3. The cumulative use time for the oral or rectal temperature probe in a single patient
should be less than 24 hours.
The monitor’s state can change from the
PREDICT
mode into the
MONITOR
mode, but it can
not change from the
MONITOR
mode into the
PREDICT
mode.
14.1.3 TEMP Setup for T2A Module
Enter
TEMP SETUP
to set the following items:
MEASURE MODE
: Set the measuring mode to
PREDICT
or
MONITOR
.
MEASURE POS
: Set the measuring position to
ORAL, AXILLARY
or
RECTAL
. The
axillary sensor can be used for measuring oral/axillary temperature, while the rectal sensor
for measuring rectal temperature.
Make sure
MEASURE POS
is properly set up every time before you withdraw the probe from
the probe well. After withdrawing the probe from the probe well, make sure
MEASURE MODE
is properly set up.
14.2 Quick TEMP with F3000 Module
14.2.1 Introduction
iM3 with the F3000 module measures patient temperatures by oral, axillary or rectal means.
The monitor can only measure temperature of adult and pediatric patients. If the user measure
temperature of neonate patient, the monitor will not display data.
The electromagnetic compatibility of this device has been verified by test according to the
EN60601-1-2: 2015 requirements.