Vital Signs Monitor User Manual Networking
- 46 -
the same to the monitor.
For detailed information, please refer to
CMS-LITE Data Management Software User Manual
.
7.7 Gateway Communication
The monitor can be connected to the Gateway (V1.21 or above), which provides clinicians with
the capability of viewing and collecting patient data remotely and the data exchange of selected
clinical and administrative information between the monitor and the hospital network.
To set the monitor server IP address, select
User Maintain
>
Network
>
Server IP
. Make sure
the monitor share the same server IP with the computer in which the MFM-CMS software is
installed.
For more information about Gateway communication, refer to
GW1
Gateway User Manual
.
7.8 HL7 Communication
The monitor supports HL7 protocol to upload data. Select
User Maintain
and input user
maintain password >
Security
. In this menu:
•
Set
HL7
to On/Off. To avoid hacker attacking, setting HL7 to Off is normally recommended.
To set HL7 IP address of the client-side, select
User Maintain
>
Network
>
HL7 IP
.
For more information about HL7 communication, refer to
HL7 Communication Protocol Service
Manual
.
7.9 Inquire for Patient Information via Network (ADT)
If the monitor is connected to network server and
ADT Query
(in
User Maintain
) is also set to
On
(default is Off), the monitor can inquire for patient information from the server.
1
In Monitor mode or Ward round mode, press Admit/Create new patient hardkey
on
the front panel to enter
New Patient
interface;
In Spot-checking mode, select
Menu
>
Patient Management > New Patient
to enter
New
Patient
interface;
2
In Monitor or Spot-checking mode, input MRN for query;
In Ward round mode, input the 1
st
patient information item, Or, directly scan barcode to get
patient information. The patient information can be updated to the monitor.
NOTE:
1
If the patient information inquired from Network server is modified on the network
server, the monitor can update the modified contents correspondingly only when the
monitor is in the monitor mode and relevant settings on the network server have also
been set to on; If patient information is modified on the monitor only in the monitor
mode, the modified contents can be updated to network server.
2
When the monitor is in the on-screen keyboard interface, scanning barcode is used
only for information input. After information input, user should complete the new