Vital Signs Monitor User Manual Alarms
- 52 -
Remind signal: Audio alarm off symbol
and
Audio Alarm off
on a red colored background
are displayed with an interval of 2 s during the audio alarm off status.
Pressing shortcut key
again can resume the audio alarm.
NOTE:
If a new alarm occurs during the audio alarm off period, the new alarm will not be
sounding.
8.3.4 Alarm Reset
In monitor mode, select the shortcut key
on the screen directly. When the alarm is
reset,
No alarms are sounding until a new alarm occurs.
As for the active alarms, the visual alarm indications are still displayed.
All latching alarms are cleared. If the alarm condition is no longer present, all alarm
indications stop and the alarm is reset.
It will not influence the configuration of physiological alarm off, audio paused, and audio off
status.
NOTE:
If a new alarm occurs after the alarm is reset, the new alarm will be sounding.
8.4 Latching Alarms
In monitor mode, to configure the alarm latching setting, select
Menu
>
Maintenance
>
User
Maintain
>
Alarm
and choose
Alarm Latch
which can be set to
On
or
Off
. When it is set to
Off
, alarm indications end when the alarm condition ends. When it is set to
On
, the visual and
audio alarm indications are still displayed after the alarm condition ends; meanwhile, the alarm
time is also displayed for the latched alarm for your reference. The indication lasts until you
acknowledge the alarm.
You can use the permanent key
on the screen to acknowledge the latched alarm.
8.5 Alarm of SpO
2
Sensor Off
In monitor mode, user can enter
Menu
>
User Maintain > Alarm
to set the alarm level
of
SpO
2
Sensor Off
to
High
,
Med.
or
Low
. It is set to
Low
by default.
In Ward round or spot-checking mode, it is acted as prompt and the level is not user-selectable.