Vital Signs Monitor User Manual Alarm Information
- 55 -
Chapter 9 Alarm Information
9.1 Physiological Alarm Information
WARNING
During monitoring, the physiological alarms
SpO
2
No Pulse
and
SpO
2
Desat
are preset
to be on and cannot be turned off.
Message
Cause
Alarm level
SpO
2
SpO
2
High
SpO
2
measuring value is above upper alarm limit.
User-selectable
SpO
2
Low
SpO
2
measuring value is below lower alarm limit.
User-selectable
SpO
2
No Pulse
The signal of the measurement site is too weak due to
insufficient blood supply and environmental factors, so
the monitor can’t detect the pulse signal.
High
SpO
2
Desat
SpO
2
measuring value is below the SpO
2
Desat Limit.
High
PR
PR High
PR measuring value is above upper alarm limit.
User-selectable
PR Low
PR measuring value is below lower alarm limit.
User-selectable
NIBP
SYS High
SYS measuring value is above upper alarm limit.
User-selectable
SYS Low
SYS measuring value is below lower alarm limit.
User-selectable
DIA High
DIA measuring value is above upper alarm limit.
User-selectable
DIA Low
DIA measuring value is below lower alarm limit.
User-selectable
MAP High
MAP measuring value is above upper alarm limit.
User-selectable
MAP Low
MAP measuring value is below lower alarm limit.
User-selectable
TEMP
TEMP High
Measuring value of TEMP is above upper alarm limit.
User-selectable
TEMP Low
Measuring value of TEMP is below lower alarm limit.
User-selectable