A
"7
Ω
load" circuit, as shown in figure 7.3, should be built in a suitable enclosure
and provided with an appropriate male AC mains plug. Note that the 1M
Ω
resistor’s stand-off voltage should be greater than or equal to the peak line
voltage.
6.8ž ±5%, 1W
0.47µF ±10%, 250VAC (X2 type)
L
N
E
open
1Mž,
±5%,
1/4W
Figure 7.3
"7
Ω
load" Circuit
•
An impedance circuit shown in figure 7.4 should be built in a suitable enclosure
with bulkhead BNC jacks (AMP 413771-x or equivalent). This impedance circuit,
placed in series with the output path of a PLCA-22 analyzer, effectively increases
its output impedance. This will allow for a more sensitive measurement of the
receive mode impedance of the product under test.
88.7ž ±1%, 1/4W
Figure 7.4
Impedance Circuit Used For Excessive Loading Verification
•
An attenuation circuit shown in figure 7.5 should be built in a suitable enclosure
with bulkhead BNC jacks (AMP 413771-x or equivalent). This attenuation
circuit, or "pad", will provide approximately 60dB of attenuation when connected
into the receive performance verification set-up of figure 7.9.
30.1kž ±1%, 1/4W
Figure 7.5
Attenuation Circuit Used For Receive Performance Verification
L
ON
W
ORKS
PLT-22 Transceiver User’s Guide
7-5
Summary of Contents for LONWORKS PLT-22
Page 6: ...iv Echelon...
Page 14: ...1 8 Introduction...
Page 67: ...LONWORKS PLT 22 Transceiver s User Guide 5 7 Figure 5 3 Capacitor Input Power Supply Schematic...
Page 92: ...6 10 Design and Test for Electromagnetic Compatibility...
Page 110: ...7 18 Communication Performance Verification...
Page 114: ...8 4 References...
Page 118: ...A 4 Appendix A...