Once the Display Backlight colour is chosen as per the choice, press
‘ENTER’ button to select, and after selection the display will come to
the previous menu.
(Note- It is available only in LCD Variants)
2.5 Zero Tracking (Ztrc)
When ENTER key is pressed from Ztrc 01, the display will start
flashing ‘01’. Default value is 01.
User can set the zero tracking value between 00 to 99 based on the
site/environment conditions.
For example if we enter 05 here and the scale least count/division is 2
g, then 2 g x 5 = 10 g, so up to 10 g display will track the values to
zero.
2.6 Zero Lock (ZLOC)
When ENTER key is pressed from ZLOC 00, the display will start
flashing ‘00’. Default value is 00.
User can set the zero lock value between 00 to 99 based on the
site/environment conditions.
For example if we enter 05 here and the scale least count/division is 2
g, then 2 g x 5 = 10 g, so up to 10 g display will show only zero and
when we add the weights more than 10 g, then it will show the actual
value kept on the scale.