Technical specifi cations
Electromagnetic compatibility (EMC)
10
11
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS
EMISSION TEST
COMPLIANCE
RF emissions – CISPR 11
Group 1
RF emissions – CISPR 11
Class B
Harmonic emissions (IEC 61000-3-2)
Class A
Voltage fl uctuations/Flicker (IEC 61000-3-3)
Compliant
Device class (in accordance with Directive 93/42/EEC
amended by Directive 2007/47/EC)
IIa
Class of electrical safety device
Class II
Type of applied parts
BF
Degree of protection
IP22
Expected life of accessories
1 year
Expected life of the device
300 hours
Capacity of the collection container
100 ml
Weight of the device
200 g without battery
Dimensions
120 mm x 90 mm x 60 mm
Voltage
Input: 100-240V (AC) – 50/60Hz – 0.5A Max.
Output: 6V (DC) – 1.0A
Length of the cable
1.8m
Absorbed power
5.2 W
Suction Values
Stimulation phase: -80 to 140 mbar
Expression phase: -155 to 280 mbar
Maximum suction: <-330 mbar
Rate
40 to 120 cycles/min.
Sound level
< 56 dB(A)
Power supply
Alternating current 100-240 V - 50/60 Hz
Battery
4 x 1.5V - AA, LR06
Room temperature (T)
for operation
+5°C < T < +40°C
Room temperature (T)
for transport and storage
-25°C < T < +70°C
Relative humidity (H) for transport,
storage and operation
15% < H < 90%
Atmospheric pressure (P) for
transport, storage and operation
700 hPa < P < 1,060 hPa
Essential performance features:
•
Provide adequate suction so that
mothers can express their milk.
•
Control the rate of suction to best
meet the baby’s sucking rhythm.
•
Equipment with moderate
compression and low fl
ow.
Intermittent suction.
Part applied to the body:
Breast shield.
NOTE: The CE declaration of
conformity and a technical description
can be supplied on demand.
Identifi cation:
Each breast pump has a batch
identifi cation that indicates the year
of manufacture.
Electromedical devices require special precautions with regard to EMC and must be installed and set to operate in accordance
with the EMC information provided in this manual.
WARNING:
Avoid using this unit near or stacked onto other equipment as this may cause a malfunction. If this is necessary,
this appliance and the other appliances should be monitored to check they are operating normally.
WARNING:
Use of accessories, transducers and cables other than those specifi ed or supplied by the manufacturer of this
device may cause an increase in electromagnetic emissions or a decrease in the immunity of this device and cause improper
functioning.
WARNING:
Portable RF communication devices (including peripherals such as aerial cables and external aerials) should
not be within 12 inches (30 cm) of any part of the MINIKIT DUO breast pump, including cables specifi ed by the manufacturer.
Otherwise, performance of these devices may be impaired.
WARNING:
If the device malfunctions due to electromagnetic disturbances, the performance of the device may be
impaired and result in delayed patient care.
Wireless communication devices such as wireless home networking devices, mobile phones, cordless phones and their bases,
walkie-talkies can aff ect the device, and it is recommended to keep them at a distance of at least 0.5 m from the device.
Summary of Contents for Kitett MINIKIT DUO
Page 3: ...00 00 00 01 L5 C 1 2 3 4 5 6 7 1 2 3 D...
Page 4: ...1 x2 5 x2 x2 x2 x2 2 3 4 6 7 8 9 10 13 14 12 11 15 x2 x2...
Page 48: ...www dtf fr MINIKIT DUO Hz V DTF MEDICAL 48 AR...
Page 49: ...MINIKIT DUO MINIKIT DUO Clean Valve MINIKIT DUO 00 15 L5 00 15 00 00 01 00 49 AR...
Page 51: ...Clean Valve 51 AR...
Page 54: ...53 53 53 53 53 52 52 52 52 51 51 50 50 50 49 49 48 48 48 48 47 47 47 46 46 45 54 AR...
Page 55: ......
Page 56: ......
Page 57: ......
Page 58: ......
Page 59: ......