portion of which is focused on the exit slit of the monochromator by a
collimating mirror. From here the beam is passed to a sample compartment
through one of the filters, which helps to eliminate unwanted second order
radiation from the diffraction grating. Upon leaving the sample compartment,
the beam is passed to the silicon photodiode detector and causes the detector
to produce an electrical signal that is displayed on the digital display.
Unpacking Instructions:
Carefully unpack the contents and check the materials against the following
packing list to ensure that you have received everything in good condition.
Packing List
Description
Quantity
Spectrophotometer
........................................................
1
Mains Lead
......................................................................
1
Glass Cuvettes…………………………………………
1 Set of 4
Quartz Cuvettes……………………………………….
1 Set of 2
Dust Cover
......................................................................
1
Operation Manual
...........................................................
1
Software Manual………………………………………
1
Software Kit (Disc 1+ USB Lead 1+ Dongle 1)…….
1
Fuse…………………………………………………….
2
Note: The printer and auto-cell holder mentioned in this manual are all optional
accessories, they do not come standard with the instrument.
Specifications:
Wavelength Range:
190-1100nm
Spectral Bandpass:
1.8nm
Wavelength Accuracy:
±
0.3nm
Wavelength Repeatability:
0.2nm
Baseline Flatness.
± 0.001A
Stray Light
:
0.05%@220nm&360nm
Photometric Range:
0-200%T,-0.3-3.0A
Noise:
0.0005A @ 500nm
Drift:
< 0.0003A/h @ 500nm
Power Requirements:
AC90-250V, 50/60Hz
Dimensions(mm)
545L
×
468D
×
245H
Light Source:
Tungsten Halogen/Deuterium
Net Weight:
18kg
Installation:
1. After carefully unpacking the contents, check the materials with the
packing list (page 4) to ensure that you have received everything in
good condition.