29
DNA/Protein
Press
【
5
】
in main menu for “DNA/Protein” (Fig 60).
【
ESC/STOP
】
to exit.
Note:The algorithm of the test refer to Appendix A please.
Fig 60
To load previous DNA results, press
(LOAD)
and select a previously stored result (.dna)
Test
1. To use a simpler or different algorithm, you can enter your own values
for f1-f4. Press
【
F1
】
to set f1-f4. Input the value by pressing the
numeric keypad (Fig 61).
Fig 61
2. Press
【
F2
】
to select test mode. ”Absorbance difference 1” is for
testing at the wavelength 260nm,280nm and 320nm (optional),and
the ”Absorbance difference 2” is for testing at the wavelength
260nm,280nm and 320nm (optional,Fig 62). Then select with/without
reference. If selected with reference (no), the A ref. will be “0” (Fig 63).
Fig 62