Basic operation
※
Blank
There is a system baseline stored in the memory of the instrument. Usually
user may not rebuild system baseline before test. Only putting the sample into
the sample light path and the reference into the Reference Light Path, the
result can be obtained. As the system baseline always get a little change after
the instrument is powered on, it is necessary for the user to rebuild the system
baseline. There are a couple of ways to rebuild the system baseline. Select
“Yes” in Fig 5 or Press
【
0
】
In Fig 73 or Press
【
F4
】
In Fig41,
Regarding Blanking, important points list below:
A. Take measure in Basic Mode
a.
Put the reference cuvette with
reference solution
into the Reference
Light Path and the sample cuvette with
reference solution
into the
Sample Light Path. Press the key
【
ZERO
】
for blanking.
Note
1. If the reference solution is too thick,
“Energy Low…”will appear
following the “Blanking…”on the screen
(Fig 8)
.If “Energy too Low…” appears
following the “Blanking…”,the t
est will be paused and
“Warning…”will appear
on the screen.(Fig 9).
2. If no automatic changer installed “cell #1” and “Max E” will disappear in
Fig8