7
【
CELL
】
Set cell position (Only available when Auto Changer used).
【
HELP
】
Reserved key for future Function Extending, not available now.
Turn on spectrophotometer
Turn on spectrophotometer by pressing the Power Switch (IO)(see Fig1). The
instrument starts to initiate and the steps are as below:
1.
The instrument will check memory first (Fig 4), please wait or press any
key to skip this step ,after positioning filter, auto-cell changer(if installed) and
D2/W lamps, the screen display as Fig 4A. 15 minutes pass or press
【
ESC
】
,
the screen display as Fig 5,Select “No” to skip to main menu( Fig 7) and select
“Yes”(recommended) to calibrate system (Fig 6).The calibrating process
include “get dark current”, “searching 656.1nm” and “check energy”. After
finish the calibration system, go to main menu too (Fig 7).
2
. If the data in memory has been lost, the instrument will directly
calibrate system without any choice for you.
3
. If no auto-cell changer installed “cell #1” will disappear in Fig7
Fig 4
Fig 4A
Fig 5