Preparation
16
Operating Ins
tructions V
amos, 1
. US ed.
Connecting the SpO2 Sensor
Connecting the SpO
2
Sensor
Selecting the sensor
The reusable Durasensor DS-100A, 57 20 072 is included in
the original delivery package – see Ordering Information,
page 49.
Other optional sensors mentioned in the tables below are
listed in the Ordering Information, page 49.
●
Only use the SpO
2
sensors described here.
Strictly follow the Instructions for Use of the sensors.
Incorrect positioning or incorrect use can cause tissue
damage.
●
Select the sensor according to the following criteria:
— Patient weight
— Agitation and movement of the patient
— Possible application position
— Perfusion of the patient
— Duration of use
The following table is intended to help in selecting the
appropriate sensor. All available sensors with their specific
characteristics are listed.
Sensor type
OXISENSOR I-20
OXISENSOR D-20
DURASENSOR
DS-100A
OXISENSOR D-25
Age group
Infants
Children
Adults
Adults
Weight of patient
1 to 20 kg
10 to 50 kg
more than 40 kg
more than 30 kg
Duration of use
Short and long-term
monitoring.
Check application
site at least every 8
hours.
Short and long-term
monitoring.
Check application
site at least every 8
hours.
Short-term
monitoring.
Change application
site at least every 4
hours.
Short and long-term
monitoring.
Check application
site at least every 8
hours.
Patient activity
Limited activity
Limited activity
Inactive patients
only
Limited activity
Preferred measuring
point
Toe
Finger
Finger
Finger
Sterilization
*
*
in unopened, undamaged packaging
Sterile pack
Sterile pack
------
Sterile pack
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS