![Dräger Vamos Operating Instructions Manual Download Page 22](http://html.mh-extra.com/html/drager/vamos/vamos_operating-instructions-manual_2526877022.webp)
Operation in Measurement Mode
22
Operating Ins
tructions V
amos, 1
. US ed.
Applying the Durasensor DS-100A
Applying the Oxisensor D-25 and D-20 (Option)
●
Regularly check the blood circulation in locations distal
from the sensor application site. Regularly check the
correct positioning of the sensor, the condition of the skin
and the correct position of the optical elements at the
application site.
Select a different measuring location if the patient’s skin is
affected.
Applying the Durasensor DS-100A
Reusable sensor supplied with the monitor and intended for
short-term monitoring of relatively calm patients with a body
weight of more than 40 kg (88 lbs).
Preferably place on the index finger or another finger. For
large or obese patients, choose the little finger.
●
Open the clip slightly and push it onto the finger. The end
of the finger must touch the end stop, and the soft pads
must rest on the nail and the fingertip. Route leads over the
top of the fingers.
●
Make sure that the clip does not pinch or cause pressure
points.
Applying the Oxisensor D-25 and D-20 (Option)
Self-adhesive sensors for short and long-term monitoring of
patients with limited movement and activity and with a body
weight ranging from 15 kg (33 lbs) to more than 50 kg
(110 lbs).
NOTE:
Long fingernails impede the application of this sensor.
Nail varnish adversely affects measurement precision.
WARNING !
Do not use self-adhesive sensors if the patient has an all-
ergic response to the adhesive strip used.
0
1
33
7
1
63
0
1
43
7
1
63
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS