Troubleshooting
39
Operating Ins
tructions V
amos, 1
. US ed.
Troubleshooting
Vamos divides alarm messages into 3 priority, identified by
exclamation marks:
Warning = !!! Top priority message
Caution = !!
Medium priority message
Advisory = !
Low priority message
The table below is intended to help identifying possible
causes of an alarm message and to assist with prompt
corrective action.
Messages are listed in alphabetical order.
Fault
Cause
Remedy
Apnea
!!!
Breathing/ventilation stopped: No breath
detected by CO
2
measurement for 30 seconds
Immediately ventilate the patient!
Check that the breathing hoses are not
disconnected
Batt. low
Option
!!
Capacity of the optional internal battery is less
than 10 % of full capacity
Connect device to mains power
Cal Error
!
Automatic gas calibration failure
Check that sample gas can flow freely from
»
« port. Route sample line without kinking.
CO
2
INOP
!
CO
2
measurement failure
Call DrägerService
Des INOP
!
Anesthetic gas measurement failure
Call DrägerService
Enf INOP
!
Anesthetic gas measurement failure
Call DrägerService
Fan
!
The internal exhaust fan is not operating
correctly.
Call DrägerService
FetCO
2
hi
!!
FetCO
2
measured value high than max. alarm
limit
Check ventilation
FetCO
2
lo
!!
FetCO
2
measured value lower than min. alarm
limit
Check ventilation
Check that the sample line (Luer Lock) is firmly
connected
FiCO
2
hi
!!
FiCO
2
measured value higher than max. alarm
limit
Replace the soda lime in the patient system.
Check that both inspiratory and expiratory valves
(valve diaphragms) are functioning correctly
FiDes lo
!!
The inspiratory measured value of the set
anesthetic gas is lower than the min. alarm limit
Check vaporizer setting
Check that the sample line (Luer Lock) is firmly
seated
FiDes hi
!!!
The inspiratory measured value of the set
anesthetic gas is higher than the max. alarm limit
Check vaporizer setting
FiEnf lo
!!
The inspiratory measured value of the set
anesthetic gas is lower than the min. alarm limit
Check vaporizer setting
Check that the sample line (Luer Lock) is firmly
connected
FiEnf hi
!!!
The inspiratory measured value of the set
anesthetic gas is higher than the max. alarm limit
Check vaporizer setting
E