Technical Data
45
Operating Ins
tructions V
amos, 1
. US ed.
Operating Characteristics
Display range for Enflurane
0 to 10 Vol.%
Precision
±(0.15 Vol.% + 15 % rel.)
Resolution
0.1 Vol.%
Display range for Sevoflurane
0 to 10.0 Vol.%
Precision
±(0.15 Vol.% + 15 % rel.)
Resolution
0.1 Vol.%
Display range for Desflurane
0 to 22 Vol.%
Precision
±(0.15 Vol.% + 15 % rel.)
Resolution
0.1 Vol.%
Warm-up phase
4 min
Drift
Zero point
within the required specifications
Measurement of functional oxygen saturation SpO
2
Oxygen saturation SpO
2
1 to 100 % SpO
2
Resolution
1 % SpO
2
Precision (adults)
in the range 70 to 100 % SpO
2
Min. ±3,5 % SpO
2
in the range 1 to 69 % SpO
2
not specified
Precision (neonates)
in the range 70 to 100 % SpO
2
Min. ±4,5 % SpO
2
in the range 1 to 69 % SpO
2
not specified
Pulse rate
20 to 250/min
Resolution
1/min
Precision
±3/min
Trigger point
<1 s
Averaging
Slow
10 to 15 seconds
Normal
5 to 7 seconds
Fast
2 to 3 seconds
Sensors
Type
Compatible with the following Nellcor sensors:
Oxisensor, Oxiband and Durasensor
Wavelengths
660 nm (red), 920 nm (infrared)
Light energy
<10 mW
Acoustic pulse signal
A tone is generated for each pulse beat.
Operating Characteristics
AC line power supply
100 to 240 V, 50/60 Hz
15 V DC output
Sound emission corresponding to free field
measurement over a reflective plane
max. 45 dB (A)