Operation in Measurement Mode
18
Operating Ins
tructions V
amos, 1
. US ed.
Switching On
Operation in Measurement Mode
Switching On
●
Switch monitor on by pressing the power switch
into »
$
$
$
$
« position.
●
This initiates start-up and test routines. Verify that:
all LEDs light up,
three tones sound,
The LEDs in the keys and the alarm LEDs light up in the fol-
lowing sequence:
1
the yellow LED in the »
« key,
2
additionally the yellow LED in the »
g
« key,
3
additionally the yellow alarm LED
4
additionally the red alarm LED
This sequence is repeated until the display first shows a pat-
tern and lights completely afterwards, then the following
text appears on the screen:
Self Test
please wait!
After the self-test, Vamos enters Standby mode.
The yellow LED in the »
« key lights up.
The following text appears on the screen:
Standby
Software
X.XX
(with X.XX being replaced by the software revision number)
●
To start Measuring mode, press »
« key. The yellow light
will go out.
If Vamos is equipped with the battery option:
— if the battery is partially discharged the symbol
»
« appears in the field for the battery status to
indicate that the battery is charged automatically.
— if the battery is completely charged the field for the battery
status is empty.
— to test if the battery is functioning
●
disconnect mains plug, »
XX % « appears in the field
for the battery status with XX being replaced by the remain-
ing capacity of the battery, reconnect mains plug.
Measurement of SpO
2
and Pulse is immediately active.
During the warm up time of the gas sensor blanks (– –) are
displayed instead of numbers. The CO
2
measurement values
are available first.The measured values will reach the specified
accuracy after a maximum time of approx. 4 minutes.
0283
7
1
63
0303
7
4
1
1
D
Vamos
1
2
3
4
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS