19. Operation and Maintenance of Neon
®
pR
101
Dr. A. Kuntze
I
Gutes Wasser mit System
Robert-Bosch-Str. 7a
40668 Meerbusch
+49 (0) 2150-7066-0 / Fax -60
www.kuntze.com
19.5.2. Software-update via SD card
With the SD card you can also install software updates. Copy the two software files you have received
from us onto the SD card (not in a subdirectory). Plug the SD card in the instruments slot, and load the
files via SYSTEM=>MEMORY=>LOAD=>SOFTWARE.
19.5.3. Test menu
During start-up, or in case of an alarm in your control center inconsistent with the measured values, the
test menu will help you test the communication between instrument and control center.
In the test menu you can operate all relays manually, and set fix values for each available mA output.
The test menu can only be used in the Manual mode. The test menu is a submenu of the menu
IN/OUT.