20. Trouble-Shooting
109
Dr. A. Kuntze
I
Gutes Wasser mit System
Robert-Bosch-Str. 7a
40668 Meerbusch
+49 (0) 2150-7066-0 / Fax -60
www.kuntze.com
Popup message
Cause
Suggested remedies
C value out of
range
C value settings e xceed the
allowed range.
Check measuring range and repeat
reference measurement.
If you calibrate and try to save the calibration results even though the C value exceeds the allowed
range, this pop-up appears. The allowed range i s wide enough to include all usual conductivity sensors.
Therefore, the message is an indication that something is wrong with either the measurement or the
reference value.
Popup message
Cause
Suggested remedies
TDS factor
out of range.
TDS factor exceeds the
allowed range.
Check measuring range and repeat
reference measurement.
If you calibrate and try to save the calibration results even though the TDS factor exceeds the allowed
range, this pop-up appears. The allowed range covers all the usual settings. Therefore, the message is
an indication that something is wrong with either the measurement or the reference value.
Popup message
Cause
Suggested remedies
SD card error
The instrument cannot use the
SD card
No SD card
Invalid format
Invalid file
SD card full
Check if an industry-standard SD card has been plugged into the slot, and that it was plugged properly.
If you want to load settings or update files, check that the files are available on the SD card directly and
not in a subfolder. Delete and restore the files, if necessary. If you want to store settings on the SD
card, check if there is enough storage space left on the card.
This message appears also if you remove the card without deactivating the logger first.
Popup message
Cause
Suggested remedies
No appli
The software files cannot be
loaded
Please renew the 2 data
files on the SD card and try again.
For updates, both files have to be stored on the SD card, not in a subfolder. Check that the files are
available, and delete and restore them if necessary.