51
English
LEVEL 1:
n
Extension (stretching)
− Maximum
knee extension:
− 10°
− Maximum
hip extension:
7°
n
Flexion (bending)
−
Maximum
knee flexion:
120°
− Maximum
hip flexion:
115°
Note!
The programmed value and the value
measured at the patient's knee may
differ slightly.
n
Speed
The speed can be adjusted in steps of 5%
between 5% and 100%.
Default setting: 50 %
n
Warm-up protocol
During warm-up, the patient will slowly
become used to the set maximum ex-
tension and flexion values, starting from
the center position.
The warm-up protocol starts in the mid-
dle between the two maximum values
set for stretching and bending. The range
of motion increases with each cycle, until
the programmed maximum values are
reached after a total of 15 cycles.
Default setting: disabled
LEVEL 2:
n
Extension
pause
Pauses occur at the set extension limit,
just before the bending movement
starts. Pauses can be set to any value
between 0 and 59 seconds in steps of 1
second, and to values between 1 and 59
minutes in steps of 1 minute.
Default setting: no pause
n
Flexion pause
Pauses occur at the set flexion limit, just
before the stretching movement starts.
Pauses can be set to any value between
0 and 59 seconds in steps of 1 second,
and to values between 1 and 59 minutes
in steps of 1 minute.
Default setting: no pause
n
Timer (therapy time)
Default setting
is
continuous operation
.
A clock symbol in the upper right-hand
corner of the display identifies the con-
tinuous mode of operation. The clock
indicates the elapsed therapy time. In
the
continuous mode
, the device must
be stopped with the
STOP
button.
However, you can also select therapy
durations of
1 to 59 minutes in steps of
1 minute
and of
1 to 24 hours in steps of
30 minutes
.
When the time has elapsed, the unit
switches
automatically
off and stops
in the position equivalent to the set
extension value +10°. In this case, a circle
replaces the clock symbol. The circle fills
as the therapy time progresses.
n
Reverse on load
(safety function)
The device automatically starts moving in
the opposite direction of the last move-
ment when the patient's resistance (load)
exceeds the set value. Adjustable levels
for reverse on load feature: 1 - 25. At
level 1, very low resistance will cause the
device to reverse; at level 25, a high resis-
tance is required to initiate the reversal.
Default setting: Level 25
Caution!
Patient hazard –
The reverse-on-load feature is only a
safety measure in the event of cramps,
spasms, locked joints and similar condi-
tions. The manufacturer cannot be held
liable for misuse of this feature.
Summary of Contents for Artromot-K1 Classic
Page 37: ...33 Deutsch ...
Page 69: ...65 English ...
Page 103: ...99 Français ...
Page 137: ...133 Español ...
Page 171: ...167 Italiano ...
Page 205: ...201 Nederlands ...
Page 210: ...206 Notes ...
Page 211: ...207 Notes ...
Page 212: ...208 Notes ...
Page 215: ...VII ...