203
12.2 Electromagnetic immunity
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
ARTROMOT®-K1
is intended for use in the electromagnetic environment specified below.
It is the responsibility of the customer or user to ensure that the
ARTROMOT®-K1
unit is
used in such an environment.
Immunity test
IEC 60601
test level
Home healthcare
Compliance
level
Home healthcare
Electromagnetic
environment – guidance
Electrostatic
discharge (ESD) to
IEC 61 000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical fast
transient/bursts to
IEC 61 000-4-4
± 2 kV
100 kHz repetition
frequency
± 2 kV
100 kHz repetition
frequency
Mains power should be that
of a typical commercial or
hospital environment.
Surges to
IEC 61 000-4-5
± 1 kV
line-to-line
voltage
± 2 kV
line-to-ground
voltage
± 1 kV
line-to-line
voltage
± 2 kV
line-to-ground
voltage
Mains power should be that
of a typical commercial or
hospital environment.
Voltage dips
To IEC 61000-4-11
0%
U
t
; 0,5 cycle
at
0°, 45°, 90°, 135°,
180°, 225°, 270°,
315°
0%
U
t
; 1 cycle
and
70%
U
t
; 2
5/30 cycles
Single phase at 0°
0%
U
t
; 0,5 cycle
at
0°, 45°, 90°, 135°,
180°, 225°, 270°,
315°
0%
U
t
; 1 cycle
and
70%
U
t
;
25/30 cycles
Single phase at 0°
Mains power should be that
of a typical commercial or
hospital environment. If the
user of the
ARTROMOT®-K1
unit requires continued
operation during power
mains interruptions, it is
recommended that the
ARTROMOT®-K1
unit
be powered from an
uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field to
IEC 61 000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristics of a typical
location in a typical commer-
cial or hospital environment.
Voltage
interuptions to
IEC 61000-4-11
0%
U
t
; 250/300
cycles
0%
U
t
; 250/300
cycles
Note:
U
t
is the a.c. mains voltage prior to application of the test level.
Summary of Contents for Artromot-K1 Classic
Page 37: ...33 Deutsch ...
Page 69: ...65 English ...
Page 103: ...99 Français ...
Page 137: ...133 Español ...
Page 171: ...167 Italiano ...
Page 205: ...201 Nederlands ...
Page 210: ...206 Notes ...
Page 211: ...207 Notes ...
Page 212: ...208 Notes ...
Page 215: ...VII ...