h
63
7. Care, Maintenance, Transport
7.1 Care
m
Warning!
Shock hazard — Unplug the device from the
power line before cleaning.
Shock hazard, equipment damage–Liquids
must not enter the device or the control
pendant. If liquids have entered into the units,
ARTROMOT® ACTIVE-K must be immediately
checked by a service technician, before it can
be reused.
− ARTROMOT® ACTIVE-K can be disinfected
by
wiping down
with a disinfectant. Thus, it
complies with the special hygiene standards
for medical technical equipment.
−
The enclosure
and removable
leg support
assemblies
can be cleaned with commonly
used disinfectants and
mild household
detergents
.
− Only use a
damp cloth
to wipe the therapy
unit down.
m
Warning!
Patient hazard, patient contamination
Before using the device on another patient, be
sure to clean and disinfect it according to the
instructions given here.
m
Caution!
Equipment damage —
− The plastic material used is not resistant to
mineral acids, formic acid, phenols, cresols,
oxidants and strong organic or inorganic
acids with a pH value below 4.
− Use only clear disinfectants to prevent
discoloration of the device.
− Do not expose the therapy unit to strong
ultraviolet radiation (sunlight) and fire.
7.2 Maintenance
(fuse replacement)
−
Check before each use
Visually inspect the device for signs of me-
chanical damage before each use.
If you detect damage or malfunctions that
may impair the safety of the patient or of the
operator, have the device repaired before
using it.
−
Technical inspections
For safety, the devices require regular
maintenance. To maintain the functional and
operational safety, check all components for
damage and loose connections at least once
a year.
These checks should be performed by per-
sons with adequate training and experience.
Damaged and worn parts must immediately
be replaced with original spare parts by
authorized staff.
m
Warning!
Patient hazard, equipment malfunction and
damage
Refer repair and maintenance to authorized
persons. Persons are authorized after training
by a specialist trained and commissioned by
the manufacturer.
ORMED GmbH will make all documents re-
quired for servicing, such as circuit diagrams,
parts lists, descriptions or calibration instruc-
tions, available to authorized experts.
The inspections can be carried out by DJO
Global Technical Service within the frame-
work of a service agreement.
The device does not require additional
regular maintenance.
Supplied by EME Services Ltd