5.3.2 General functions
New Patient
With this function, the carriage will move to the home position (at 40° flexion),
allowing the mechanical settings to be performed. All values stored on the chip
card will be deleted.
Press the OK button to select the function, press the arrow up button to activate
the function and then press the OK button to confirm your selection. The main
menu will be displayed again. Then press the START button: the carriage enters
the home position and existing therapy parameters will be deleted.
Next, you will see the main menu where you can select the operating mode.
The "new patient" function (home position) selects the following settings:
– extension:
– flexion:
– speed:
– extension pause:
– flexion pause:
– timer:
– reverse on load:
– active zone:
– extension force:
– flexion force:
– extension force range:
– flexion force range:
– therapy protocols:
– special functions:
– operating mode:
10°
60°
50 %
0
0
disabled
60 kg
0° to 90°
+ 5 kg
- 5 kg
+ 1 kg to + 5 kg
- 1 kg to – 5 kg
Level 1
disabled
no operating mode (CPM/CAM) selected