In this situation, you may attempt to restart the
unit by turning it briefly off and on again with
the power switch. With the unit switched off,
check that all plugs are correctly connected.
If the error message persists when the unit is
switched on again.
4.2 Connecting the external muscle
stimulator
m
Caution!
Patient hazard, equipment malfunction —
Only the
ARTROSTIM® FOCUS®plus
muscle
stimulator with EN 60601-1 approval, from
ORMED GmbH (manufacturer: Empi, Inc.,
USA) may be connected to the
ARTROMOT®
ACTIVE-K
physiotherapy unit.
Use the "muscle stimulation" cable (part no.
2.0037.024) to connect the muscle stimulator.
Other cables are not approved. The cable can
be ordered separately from DJO Global.
1. Plug the jack into the
ARTROSTIM® FOCUS®
plus
muscle stimulator.
2. Connect the round plug to connector (16) of
the ARTROMOT® ACTIVE-K and turn it a little
clockwise to lock.
3. Turn on the ARTROMOT® ACTIVE-K first, then
turn on the muscle stimulator.
Note
– For information on connecting and pro-
gramming the muscle stimulator, refer to
the operation manual that comes with the
ARTROSTIM® FOCUS®plus
.
– The EMS control function is only available
in the "passive" operating mode.
– The
ARTROMOT® ACTIVE-K
only synchro-
nizes motion therapy and electrotherapy;
the therapy parameters must be pro-
grammed at each device.
4.3 Adjusting the unit to the femur
length
1. Measure the
length of the patient's thigh
(femur)
from the greater trochanter to the
outer knee joint gap (Fig. A).
Fig. A
2. Set the carriage to the home position
(see 4.1).
3. Adjust the measured value at the
femur
scale
(3) of the carriage (Fig. B):
Fig. B
• Loosen the two fixation screws (4). (Open
the screws as far as necessary to enable
adjusting the slide without any resistance.)
• Extend the scale (3) to the required length.
• Tighten the fixation screws (4) to set the
scale to the new length.
m
Caution!
Equipment damage — Do not attempt to
extend the femur scale beyond the stop.