Lock-out function
With the lock-out function, access to individual operating modes or functions
can be blocked. This reduces the scope of programming options and makes
operation easier.
To activate the lock-out feature, simultaneously press the arrow left and arrow
right buttons for approx. 4 seconds. The carriage must be in the stop mode of
the operating mode concerned (no active menu). The display automatically
shows the lock-out menu.
Five different lock-out levels are available:
Lock-
out
level
Symbol
Description
0
The lock-out feature is deactivated.
All features are available and can be selected.
1
Level 1 of the lock-out feature is enabled.
This means that the selected operating mode is
fixed and no other operating mode can be se-
lected. However, all parameters of this operating
mode can be selected and modified..
2
Level 2 of the lock-out feature is enabled.
Same as lock-out level 1. Additionally, it is not
possible to switch between the different proto-
cols of the coordination and therapy protocol
modes, such as between "ACL" and "Cartilage"
or between "Find angle, active" and "Find angle,
passive". However, all parameters of the operat-
ing mode/protocol can be selected and modified.
3
Level 3 of the lock-out feature is enabled.
Same as lock-out level 2.
Additionally, all programming features are locked
and only the range of motion can be modified.
4
Level 4 of the lock-out feature is enabled.
The menu for modification of the treatment
parameters cannot be accessed. Only the START
and STOP buttons can be used. All other buttons
are locked. The treatment parameters cannot be
modified. However, in the passive mode all but-
tons can be used as "Emergency stop" buttons.
If the lock-out function is activated, the lock-out status and the locked level are
indicated in the header during operation.