61
Eng
lis
h
Patients with
a programmed chip card
• First complete the mechanical
adjustments.
• Then insert the chip card (the patient is
not yet positioned on the CPM device).
• Press the
START
key: the device will
move to the starting position as speci-
fied by the parameters stored on the
chip card and stops.
• Position the patient on the CPM device
and press the
START
key to initiate
therapy.
5.2 Programming
ARTROMOT®-S3 Standard
Different programming levels are provided
to program the ARTROMOT®-S3 Standard
models.
You change between levels by pressing the
MENU key
repeatedly.
The display always indicates the currently
selected level.
The following
treatment values, settings
and
information
can be entered/viewed
on the programming unit (1):
LEVEL 1:
– Abduction
– Adduction
– Internal rotation
– External rotation
LEVEL 2:
– Pause
– Therapy timer
– Speed
– New patient
LEVEL 3:
– Reverse on load - motor A
– Reverse on load - motor B
– Motor A ON/OFF
– Motor B ON/OFF
LEVEL 4:
– Transport setting
– Synchronized/
non-synchronized mode
– Total therapy time
– Service menu
Note!
•
While you adjust the values,
the
carriage will move to the set range.
This allows you to easily and quickly
determine the ROM where the pa-
tient does not experience pain.
•
Special functions
can be pro-
grammed and retrieved with
ARTROMOT®-S3 Comfort
devices
(see sections 5.4 and 5.5).
• The angle entered last for the respec
-
tive direction of motion will be saved
in each case.
5.3 Treatment value details –
Standard models
• To access the programming level you
need, press the
MENU key
repeatedly.
• The treatment parameters are selected
with the different
parameter keys
.
• To change the treatment values, press
the
+ / – keys
.
• To enable/disable a function, press the
respective
parameter key
again.
• To save the entered values, press the
STOP key
.
Summary of Contents for 80.00.023
Page 135: ...131 Français ...
Page 181: ...177 Español ...
Page 225: ...221 Italiano ...
Page 270: ...266 ...
Page 273: ...VII Notes ...