53
Eng
lis
h
Definitions
Read the safety statements before use of
the CPM device. The safety statements are
classified as follows:
Danger!
This term indicates an imminent hazard.
If not avoided, this hazard will result in
death or serious injury.
Warning!
This term indicates a hazard. If not
avoided, this hazard can result in death
or serious injury.
Caution!
Indicates a potential hazard. If not
avoided, this hazard can result in minor
personal injury and/or product/property
damage.
Safety information
Danger!
Explosion hazard –
ARTROMOT®-S3
is not designed for use
in areas of medical locations where an
explosion hazard may occur. An explo-
sion hazard may result from the use of
flammable anesthetics, skin cleansing
agents and disinfectants.
Warning!
Patient hazard –
− Only
authorized individuals
are
allowed to operate the
ARTROMOT®-S3
device. Individuals
are authorized after receiving
training in the operation of the device
and reading this operation manual.
− Before using the device, the operator
must ascertain that it is in correct
working order and operating condi-
tion. In particular, the cables and
connectors must be checked for signs
of damage. Damaged parts must be
replaced immediately, before use.
−
Before therapy
, a
test run
consist-
ing of several exercise cycles must
be completed, first without and then
with the patient. Check that all set-
ting screws are tightened.
− Stop therapy immediately, when you
have doubts about the device set-
tings and/or the therapy protocol.
− Ensure an
anatomically correct
setup of the CPM device
suitable
for the patient to be treated. For
this purpose, check the following
settings/positions (see numbers on
device):
1. Horizontal adduction/abduction
2. Height adjustment
3. Adjustment of upper arm length
_________________
¹ Only with
ARTROMOT®-S3 comfort
3. Safety information
Summary of Contents for 80.00.023
Page 135: ...131 Français ...
Page 181: ...177 Español ...
Page 225: ...221 Italiano ...
Page 270: ...266 ...