Routine and Preventive Maintenance
__________________________________________________________The Electrospray and APCI Probes
APCI Probe Disassembly
The starting point for this procedure is the APCI probe removed from the
instrument.
1.
2.
3.
4.
Using the 2.5 mm Allen key, unscrew the M3 x 10 cap head stainless
steel screw (P/N FM103046).
Remove the probe clamp and locking plate.
Using the 10 mm spanner, unscrew the capillary retaining nut.
Remove the graphite ferrule (P/N 6070119), PEEK insert (P/N
FM102591), and APCI capillary tube (P/N FM102594).
Caution.
Exercise care when handling the APCI probe capillary because it
is fragile and may be damaged easily.
5.
6.
Using the 2.5 mm Allen key, unscrew the two M3 x 8 cap head stainless
steel screws (P/N 5313020).
Remove the APCI probe mount and O-rings.
Cleaning the APCI Probe
The APCI probe can be stripped down for thorough cleaning after prolonged
use with additives.
1. Wipe the surface of the stainless steel insert capillary with [50:50]
methanol:water.