Routine and Preventive Maintenance
__________________________________________________________The Electrospray and APCI Probes
ESI Probe Disassembly
The starting point for this procedure is the ESI probe removed from the
instrument.
1.
2.
3.
4.
Using the 2.5 mm Allen key, unscrew the M3 x 10 cap head stainless
steel screw (P/N FM103046).
Remove the probe clamp and locking plate.
Using the 10 mm spanner, unscrew the capillary retaining nut.
Remove the graphite ferrule (P/N 6070119), PEEK insert (P/N
FM102591), and ESI probe capillary (P/N FM102698).
Caution.
Exercise care when handling the ESI probe capillary because it is
fragile and may be damaged easily.
5.
6.
7.
Using the 2.5 mm Allen key, unscrew the two M3 x 8 cap head stainless
steel screws (P/N 5313020).
Remove the ESI probe mount and O-rings.
Remove the ESI ceramic sleeve (P/N FM103394) from the ESI ceramic
probe assembly.
____________________________MSQ Hardware Manual ____________________________
5-5