Reference Inlet System
___________________________________________________________ Standard Mass Scale Calibration
Standard Mass Scale Calibration
If there is a significant change in the observed masses, you may wish to
perform a standard mass scale calibration.
1. Ensure that there is calibrant in the calibration reference bottle. If not,
unscrew the calibration reference bottle (P/N FM102771), fill it with
calibrant, and screw it back into place.
Figure 4-2. Calibration reference bottle
2. Check the waste bottle (P/N FM102770). If it is full, unscrew it, empty
it, and then screw it back into place.
Figure 4-2. Waste bottle
3.
4.
5.
Check that liquid is emerging into the waste bottle.
Ideally, the instrument should already be in operate with the gas flow
on. If it is not, the software will turn on the gas and put the instrument
into operate.
Click, with the right mouse button, on the server icon and select
Instrument Tune and Calibration…
from the menu displayed. The
Instrument Tuning and Calibration dialog box is displayed.
____________________________MSQ Hardware Manual ____________________________
4-5