Service/maintenance
OPERATION MANUAL
Chapter 7
Water Cooler
Errors and omissions excepted.
V1.1.0en/13.12.12
48
Leave the
>Drain<
[8] open for a while to allow the cleaning agent to evaporate in the temperature
control unit.
PROCEDURE
After drying out, remove the drain hose and re-fit the knurled screw to the
>Drain<
[8].
Remove the collection vessel and dispose of the contents in a proper manner.
Clean your external open application.
Re-connect your application.
Fill the temperature control unit with thermal fluid, as described on page 36 in section
and venting externally open application«
.
Vent the temperature control unit as described on page 36 in section
. An externally open application does not need to be vented.
Restart the temperature control unit in normal mode.
7.4
Cleaning the surfaces
Exposed plug contacts
DAMAGE CAUSED BY FLUID INGRESS
Protect unused plug contacts with the protective caps supplied.
Clean surfaces damp.
H22
A standard stainless steel cleaning agent is suitable for cleaning the stainless steel surfaces. Carefully
clean paint surfaces (damp) using the suds of a mild detergent.
7.5
Inspect the mechanical seal
No visual inspection of the collecting container
DAMAGE INSIDE THE TEMPERATURE CONTROL UNIT CAUSED BY OVERFLOW FROM THE COLLECT-
ING CONTAINER
Inspect the collecting container on a monthly basis and drain as required.
H23
Because mechanical seals are never completely tight, drips can be expected to form on the me-
chanical seal when using thermal fluids that barely evaporate. These drips are collected under con-
trol. The collecting container must be regularly checked and drained as required (see page 45, sec-
tion
»Function check and visual inspection«
).
7.6
Plug contacts
Exposed plug contacts
DAMAGE CAUSED BY FLUID INGRESS
Protect unused plug contacts with the protective caps supplied.
Clean surfaces damp.
H22
Protective caps are supplied for all plug contacts. Make sure that any plug contacts not required are
protective with the caps.
INFORMATION
.
NOTE
.
NOTE
.
NOTE
.
Summary of Contents for BioAcc-cool
Page 1: ...Cat No BioAcc cool Water Cooler Version 1 1 ...
Page 6: ...OPERATION MANUAL Water Cooler Errors and omissions excepted V1 1 0en 13 12 12 6 ...
Page 10: ...OPERATION MANUAL Water Cooler Errors and omissions excepted V1 1 0en 13 12 12 10 ...
Page 58: ......
Page 59: ...PAGE 55 www diagenode com ...
Page 60: ...orders diagenode com orders na diagenode com www diagenode com ...