Service/maintenance
Chapter 7
OPERATION MANUAL
V1.1.0en/13.12.12
Errors and omissions excepted.
Water Cooler
45
7.2
Maintenance
Cleaning/maintenance while the temperature control unit is operating
MORTAL DANGER FROM ELECTRIC SHOCK
Stop an ongoing temperature control process.
Disconnect the temperature control unit from the power supply by turning the
>Mains switch<
[37] on the temperature control unit to “0”.
Also disconnect the temperature control unit from the current supply.
G06
Carrying out maintenance work not described in this operation manual
DAMAGE TO THE TEMPERATURE CONTROL UNIT
For maintenance work not described in the operation manual, contact Peter Huber
Kaeltemaschinenbau GmbH.
Maintenance work not described in this operation manual is reserved for qualified specialists
trained by Huber.
Carry out only the following maintenance activities on the temperature control unit.
H16
7.2.1
Function check and visual inspection
Cooling* Description
Maintenance interval Comment
Person responsible
A
Visually inspect hoses
and hose connections
Prior to switching on
the temperature
control unit
Replace leaking hoses and
hose connections prior to
switching on the
temperature control unit.
User and/or
operators
A
Inspection in
accordance with
the F-Gas Directive
In accordance with
the F-Gas Directive
See page 15, section
»Temperature control units
with 3 to 30 kg fluorinated
greenhouse gases/natural
User
A
Inspect power supply
cable
Prior to switching on
the temperature
control unit or on
relocation
Do not start-up the
temperature control unit
if the power cable is
damaged.
Qualified electrician
(BGV A3)
A
Clean air inlet grille
As required
Clean air inlet grille with a
damp cloth.
User
A
Thermal fluid
inspection
As required
User and/or
operators
A
Changing the thermal
fluid
Weekly
See page 47, section
User and/or
operators
A
Cleaning the thermal
fluid circuit
Every 14 days
See page 47, section
User and/or
operators
A
Inspect mechanical
seals (collecting
container)
Monthly
See page 48, section
User and/or
operators
A
Check condenser fins
As required, after
3 months at the latest
See page 46, section
»Clean condenser fins (air-
cooled temperature control
User and/or
operators
A
Inspect temperature
control unit for
damage and stability
Every 12 months or
after relocation
User and/or
operators
*A = Air cooling
HAZARD
NOTE
.
Monitoring interval
Summary of Contents for BioAcc-cool
Page 1: ...Cat No BioAcc cool Water Cooler Version 1 1 ...
Page 6: ...OPERATION MANUAL Water Cooler Errors and omissions excepted V1 1 0en 13 12 12 6 ...
Page 10: ...OPERATION MANUAL Water Cooler Errors and omissions excepted V1 1 0en 13 12 12 10 ...
Page 58: ......
Page 59: ...PAGE 55 www diagenode com ...
Page 60: ...orders diagenode com orders na diagenode com www diagenode com ...