Commissioning
Chapter 2
OPERATION MANUAL
V1.1.0en/13.12.12
Errors and omissions excepted.
Water Cooler
23
Clearance to the temperature control unit in cm
Side of the temperature
control unit
Temperature control unit
with air cooling
Temperature control unit
with water cooling
Top
Air outlet on top of unit: free
standing; otherwise can be
located under a bench
Can be located under a bench
Front
min. 20
min. 10
Right
min. 20
min. 10
Left
min. 20
min. 10
Rear
min. 20
min. 20
Clearance to the temperature control unit in cm
(for operation in a bath)
Top
Air outlet on top of unit: free
standing; otherwise can be
located under a bench
Can be located under a bench
Front
min. 20
min. 20
Right
min. 20
min. 20
Left
min. 20
min. 20
Rear
min. 20
min. 20
Ambient temperature values are provided on the technical data sheet; to ensure trouble-free op-
eration, compliance with the ambient conditions is mandatory.
Maximum relative humidity 80% to 32 °C and to 40 °C decreasing linearly to 50%.
Short distance to supply connections.
The temperature control unit must not be installed so as to hinder or prevent access to the isola-
tor (to the power supply).
Magnitude of the power supply fluctuations: see data sheet from page 55 in section
Transient surges, as would normally occur in the power supply system
Applicable degree of soiling: 2.
Surge category II.
Observe page 18 of section
»Schematic illustration of the cooling variants«
.
Summary of Contents for BioAcc-cool
Page 1: ...Cat No BioAcc cool Water Cooler Version 1 1 ...
Page 6: ...OPERATION MANUAL Water Cooler Errors and omissions excepted V1 1 0en 13 12 12 6 ...
Page 10: ...OPERATION MANUAL Water Cooler Errors and omissions excepted V1 1 0en 13 12 12 10 ...
Page 58: ......
Page 59: ...PAGE 55 www diagenode com ...
Page 60: ...orders diagenode com orders na diagenode com www diagenode com ...