3006.0054.99
Order-No.:
Water Cooler
Technical data according to DIN 12876
Operating temperature range
-20...40 °C
Temperature adjustment
digital
Temperature indication
digital
Internal temperature sensor
Pt100
Resolution of display
0,1 K
Temperature stability at -10°C
0,2 K
Safety classification
Class I / NFL
Cooling power
at 15°C
0,3 kW
at 0°C
0,2 kW
at -10°C
0,14 kW
at -20°C
0,07 kW
Refrigeration machine
air-cooled, natural
refrigerant
Refrigerant
R290
Refrigerant quantity
0,04 kg
Circulation pump
Pressure- and suction
pump
max. delivery
20 l/min
max. delivery pressure
0.2 bar
max. delivery (suction)
17 l/min
max. delivery pressure (suction)
0,18 bar
Pump connection
G 1/4" female thread
min. filling capacity
1,4 l
Volume of expansion
2,6 l
Overall dimensions WxDxH **
225x360x380 mm
Net weight
23 kg
Power supply requirement
230V 1~ 50/60Hz
max. current
2,8 A
Fuse (1 phase)
4 A
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
170661
1.0/12
Technical details and dimensions are subject to change. No liability is accepted for errors or omissions.
Peter Huber Kältemaschinenbau GmbH Werner-von-Siemens-Str. 1 D-77656 Offenburg Tel 0781/9603-0 Fax 0781/57211 www.huber-online.com
Chiller with air-cooled refrigerating unit and circulation pump. Evaporator (cooler), tank and housing of stainless steel.
Pressure-suction pump made of industrial plastic material. Digital Temperature adjustment and digital temperature display.
Level indicator with sight glass.
Special Case: Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the
manufacturer). It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection. A more
resistant plastic is available on request at an additional cost.
MPC-Controller:
Modern and easy to use microprocessor controller with a large temperature display.
Limited to essential functions only:
* Large temperature display
* LED indicators for pump, cooling and heating
* Simple operation using only 3 keys
Accessories and periphery: Hose coupling nom. dia. 8*/12*, sleeve nuts thread M16x1*, cover expansion tank*, drain
valve, connection tubes
* standard equipment
Output data valid for: Room temperature 20°C
in accordance with EN60034-1 the following voltage and frequency tolerances are valid:
V / - 5% with a simultaneous frequency tolerance of + / - 2%
Example
-5% voltage and +2% frequency -> not allowed!
-5%voltage and -2% frequency -> allowed
.
** Please respect space requirements. See operating conditions at
www.huber-online.com
Summary of Contents for BioAcc-cool
Page 1: ...Cat No BioAcc cool Water Cooler Version 1 1 ...
Page 6: ...OPERATION MANUAL Water Cooler Errors and omissions excepted V1 1 0en 13 12 12 6 ...
Page 10: ...OPERATION MANUAL Water Cooler Errors and omissions excepted V1 1 0en 13 12 12 10 ...
Page 58: ......
Page 59: ...PAGE 55 www diagenode com ...
Page 60: ...orders diagenode com orders na diagenode com www diagenode com ...