7 Maintenance and cleaning
Dentsply Sirona
7.7 Coolant system
Operating Instructions inLab MC X5
7.7
Coolant system
The machine is equipped with a coolant circuit for wet processing. This
must be checked and maintained regularly. The coolant circuit includes
the coolant nozzles and the tank system.
7.7.1
Coolant nozzles
The machine is equipped with three coolant nozzles for cooling the tool.
Check the nozzles monthly for proper operation. To do this, proceed as
follows:
1. Remove the suction bellows from the mandrel (see "Suction bellows
2. Start the rinsing process in the machine configuration menu in the
inLab software on the machine.
3. Check the coolant jets impinge on the tool in a properly bundled way.
4. Next remove any foreign bodies in the coolant nozzles carefully using
a probe. Use a spray (if available) and a cleaning wire.
5. Mount the suction bellows on the mandrel (see "Suction bellows
CAUTION
Risk of injury
There is risk of injury through sharp-edged restorations as well as sharp-
edged material residues in the tank.
➢ Pay attention to the sharp-edged restorations / material residues
after processing, during cleaning and coolant changes.
NOTICE
Check coolant level
Before every wet processing check the coolant level using the
min /
max
marking. Top up the coolant if necessary.