![DENTSPLY inLab MC X5 Operating Instructions Manual Download Page 18](http://html1.mh-extra.com/html/dentsply/inlab-mc-x5/inlab-mc-x5_operating-instructions-manual_2483578018.webp)
64 78 759 D3608
18
D3608
.
201.01.07
.
02 10.2017
2 General information on the device
Dentsply Sirona
2.5 Technical data
Operating Instructions inLab MC X5
2.5
Technical data
Type designation
inLab MC X5
Rated line voltage
AC 100V ... 240V
Rated power frequency
50/60 Hz
Nominal current
2.0A
Nominal power output
200VA
Permissible line voltage
fluctuations
±10% of nominal voltage
Compressed air connection
At least 7.0bar / 0.7MPa / 102psi
upstream of filter regulator input
Filter regulator setting
6.5 - 7.0bar
Compressed air consumption
At least 50 l/min (1.8cfm)
Briefly up to 100 l/min (3.6cfm)
Recommended for dry operations:
At least 80 l/min (2.8cfm)
Type of protection against electric
shock
Class I device
Degree of protection against
ingress of water
Usual device (without protection
against ingress of water), IP 20
Overvoltage category
II
Shock stress
1J (IK06) on door glass,
5J (IK08) on remaining device
The impact strength on the door
glass was tested with a pendulum
blow test from a vertical falling
height of 400 mm.
Ambient conditions
For indoor use
Pollution degree 2
Air pressure: 700 hPa – 1060 hPa
Operating height: ≤ 3000 m above
sea level
Temperature range
+5°C up to +40°C
(+41°F - 104°F)
Humidity range
80% rel. up to 31°C (88°F)
decreasing to
50% rel. up to 40°C (104°F)
Operating mode
Continuous operation
Dimensions W x H x D
in mm
in inches
629 x 795 x 575
24 ¾ x 31 ⅜ x 22 ⅝
Approx. weight
87kg (191.8lbs)