![DENTSPLY inLab MC X5 Operating Instructions Manual Download Page 48](http://html1.mh-extra.com/html/dentsply/inlab-mc-x5/inlab-mc-x5_operating-instructions-manual_2483578048.webp)
64 78 759 D3608
48
D3608
.
201.01.07
.
02 10.2017
5 Commissioning
Dentsply Sirona
5.2 Putting the unit into operation
Operating Instructions inLab MC X5
Automatic IP settings
There is an option for integrating the machine into an existing network
with DHCP server in such a way that the IP address is received by the
DHCP server automatically. To do this, proceed as follows:
1. Establish a direct connection with the inLab 4 PC (see Installing the
2. If you want to change the IP address to automatic addressing, click
"Edit Device Settings"
.
3. Now click
"Auto IP settings"
.
4. Confirm the new network settings using
"Ok"
or press
"Cancel"
in
order not to save the changed settings.
5. If you have changed the setting to
"Automatic"
you must now
disconnect the Ethernet connection to the PC and connect the
machine to the network socket in your local network.
6. If you wish to change the settings back again then change over to
"Manual IP settings"
.
Suction
You can define the type of suction used using the
"Suction"
setting.
●
"Sirona"
: You are using the optional suction available for inLab MC
X5. Select this option and connect the suction to the machine
(including interface cable). The communication between inLab MC
X5 for starting and stopping the suction etc. is then automatic.
●
"Other"
: You are using a different, external suction. Select this option
and ensure that the requirements for external suction (see Suction
requirements [ → 28]) are fulfilled. The suction must then be switched
on and off manually for each process.
Use Spindle Touch
Using the
"Use Spindle Touch"
setting, you can choose whether the
selected block size should be checked, and if the exact position of the
block in the machine should be determined, for the inLab MC X5 during
grinding and milling processes.
IMPORTANT
Ensure that the network in which you are integrating the machine has an
active DCHP server for the allocation of IP addresses.
IMPORTANT
If the machine is in
"Auto IP settings"
mode and does not detect any
active DHCP server, the IP is automatically reset to 192.168.230.1. You
can then add and manage the machine again using a direct connection
to a PC via the
"Add Device (Manual)"
function.
IMPORTANT
Please note that in cases where this option (
"Use Spindle Touch"
) is not
activated, grinding processing of the meso structure blocks is not
possible.