6 Operation
Dentsply Sirona
6.2 Machining processes
Operating Instructions inLab MC X5
6. Now slightly tighten the screws on the fixing ring with the help of the
torque wrench for the tool holder in an alternating manner. Make sure
that the fixing ring is parallel to the base and is not slanted. If the ring
is in parallel, firstly tighten the three screws lightly (not with full torque)
and then finally tighten the screws in succession using the torque
wrench and the pre-selected torque. The required torque is reached
if a clicking/ popping sound is heard when tightening the screws. Use
only the torque wrench supplied with the workpiece holder from
Sirona for fixing the workpiece.
7. Close the door of the machine after inserting and fixing the
workpiece, and follow the instructions of the software.
6.2.5
Loading with the multi-block holder
During each change, ensure clean contact surfaces of all parts;
otherwise, the fixation and the correct positioning cannot be guaranteed.
Depending on your choice of materials, also follow the processing
instructions of the manufacturer.
For the processing of material blocks, the workpiece holder can be
equipped with one multi-block holder and with different blocks (see
"Equipping the multi-block holder [ → 61]“).
In order to load the workpiece holder with the multi-block holder,
proceed as follows:
1. Open the door of the machine by pressing the Unlock door button.
ª
The mandrel and the workpiece holder have automatically moved
to the loading position beforehand.
CAUTION
Risk of injury while reaching out to the processing chamber
There is the risk of cut injuries and crushing through sharp tools and
moving parts.
➢ Make sure that you do not touch the sharp tools and moving parts.