5 Commissioning
Dentsply Sirona
5.2 Putting the unit into operation
Operating Instructions inLab MC X5
5.2.7.2.1.3
Service menu (only inLab MC X5)
You can do the following in the service menu:
● Spindle maintenance
● Moving to the cleaning position
● Switching the water pump on/off
● Switching the suction on/off
● Opening the chuck
Spindle maintenance
1. Click on the start button to start the spindle maintenance.
2. Then simply proceed as prompted by the software.
3. Read also which steps are to be performed in the inLab MC X5
operating instructions under the section "Tool spindle maintenance".
Moving to the cleaning position
➢ Click on the start button to move to the cleaning position.
ª
The machine moves to the cleaning position.
Starting the cleaning process
➢ Click on the start button to start the automatic cleaning cycle.
ª
The machine moves for cleaning essential areas with the spindle
using the coolant.
Switching the water pump on/off
1. Click on the start button to turn on the water pump.
ª
The water pump is switched on.
2. Click on the pause button to turn off the water pump.
ª
The water pump is switched off.
Opening the chuck
➢ Click on
"Open chuck"
to be able to manually take instruments out of
the chuck.
IMPORTANT
If the clamped instrument cannot be removed from the machine
automatically, you can replace it manually through
"Spindle
Maintenance "
. To do this click on the start button for the spindle
maintenance and follow the on-screen instructions.
IMPORTANT
Then ensure that the chuck is properly tightened and the dummy tool is
placed in its position.