7 Maintenance and cleaning
Operating Instructions inLab MC X5
7.4 Maintenance for the tool mandrel
båÖäáëÜ
7.4.4
Tool has broken or not been stored
If a tool breaks during the machining process, it can no longer be stored
automatically; proceed as follows:
1. Access the service mode for maintenance of the collet chuck in the
software and continue to follow the instructions of the software.
2. Remove the collet chuck, as described in the Section "Cleaning and
maintenance of the collet chuck / collet chuck change [ → 74]".
3. Remove the remainders of the broken tool from the collet chuck.
4. Clean the collet chuck and insert the collet chuck back, as described
in Section "Cleaning and maintenance of the collet chuck / collet
5. Continue to follow the instructions of the software.
CAUTION
Risk of injury
A broken tool can be hot and sharp-edged.
➢ Wait until the tool is cooled.
➢ Use gloves to protect against injury.