7 Maintenance and cleaning
Operating Instructions inLab MC X5
7.6 Filter regulator
båÖäáëÜ
7.5.3
Processing chamber screen
The machine is equipped with a processing chamber screen (B) for
collecting material residues. Check the screen regularly.
If there are many chips and machining residues on the screen of the
processing chamber, remove it, and empty it.
7.6
Filter regulator
The machine is equipped with a filter regulator, which ensures, among
other things, the air quality in the system.
Check the filter regulator weekly for proper operation.
1. Check if condensate has accumulated in the separation vessel (A).
2. If condensate has accumulated here, drain it by pressing the drain
plug (B).
3. Then, make sure that the drain plug is locked again properly.
4. Check that 7bar / 0.7MPa / 102 psi pressure is in contact on the filter
regulator.