6 Maintenance
6.4 Replacing the O-ring
67 97 216 D3801
D3801.201.01.03.02 2022-09
71
1.
Slide the O-ring at one point into the O-ring groove.
2.
Starting from this point, carefully use your thumb and index finger to
push the O-ring into the groove (in a similar way to pushing it out
but without any force).
3.
Guide your thumb and index finger in such a way along the groove
until the O-ring is fully positioned in the groove.
NOTE
The O-ring can be damaged
Ensure that the O-ring fits in the groove without twisting in order to
prevent damage to the O-ring when fitting the sleeve.
IMPORTANT
Replacement O-rings
Replacement O-rings are available with the scanner or can be
ordered under REF 66 80 974.