6 Maintenance
6.1 Cleaning, disinfection, and sterilization
Dentsply Sirona
Operating Instructions Primescan Connect
56
67 97 216 D3801
D3801.201.01.03.02 2022-09
9.
Insert the metal bracket with the sleeve.
10. Place the cover on the container and leave the sleeve in the
disinfectant for at least 12 minutes (CIDEX
®
OPA). The sleeve will
not be damaged if it remains in the disinfectant for longer than the
required time, but removing it shortly after the required time is
recommended.