5 Operation
5.2 Scanner guide
Dentsply Sirona
Operating Instructions Primescan Connect
38
67 97 216 D3801
D3801.201.01.03.02 2022-09
5.2
Scanner guide
CAUTION
After each use
Prepare the scanner once again after each patient.
➢ Follow the instructions in the "Scanner [→ 49]" section on
cleaning, disinfection and sterilization in order to avoid cross-
contamination between patients.
The scanner acquires images which are used during the ongoing
measurement in spatial relation to each other (image registration).
During the acquisition and then during the ongoing registration process,
a distinctive sound can be heard.
If the registration cannot be implemented, the acquisition flow is
suspended. You are informed of this by means of a sound. This is
different to the sound emitted during successful acquisition. You can
adjust the volume under configuration and select another type of sound
(melody).
IMPORTANT
Registration error
Should a registration error occur, you must return to another acquired
point.
To start with, practice this procedure on the model and then on
intraoral areas.
➢ Move the scanner to a position where a successful acquisition
was taken. A point that has already been acquired in the occlusal
area is best.
Ä
You will be able to hear the sound for registered acquisitions.
➢ Continue the acquisition.
Divide the acquisition into four consecutive sequences:
1.
Occlusal
2.
Buccal
3.
Lingual
4.
Proximal