Operating Instructions Primescan Connect
6 Maintenance
6.1 Cleaning, disinfection, and sterilization
67 97 216 D3801
D3801.201.01.03.02 2022-09
49
6.1.3
Scanner
CAUTION
Risk of injury
An evidently damaged scanner must no longer be used on patients.
If the Primescan scanner accidentally falls off, check to make sure
that the sapphire glass is not damaged. If the Primescan scanner has
been damaged, it must no longer be used on patients.
The Primescan scanner must be recalibrated.
NOTE
Do not sterilize the Primescan scanner and the scanner cable!
NOTE
The window sleeves cannot be sterilized in the autoclave.
NOTE
For markets where the RKI* guidelines are to be observed
The window sleeve is classified as a "semicritical medical device A"
according to RKI guidelines and therefore does not have to be
autoclaveable.
*RKI=Robert Koch Institute, Berlin (Germany).