2 Safety instructions
2.1 Basic safety information
67 97 216 D3801
D3801.201.01.03.02 2022-09
13
WARNING
Danger of touching live parts
If the housing is damaged, there is a possibility of touching live parts
inside the unit.
➢ Check that the unit is intact. The unit can only be used for work
purposes if it is intact.
➢ If the housing is damaged, the unit must be put and left out of
operation until it has been professionally repaired.
CAUTION
Restoration to be checked by trained personnel
Each restoration created must be checked for suitability by a trained
person (e.g. dentist).
IMPORTANT
Immediately notify the manufacturer and the national competent
authority of any serious incident that has occurred with the product.
Plug connections of external PC interfaces
CAUTION
Additional devices connected to external interfaces must be tested
according to the relevant standards, e.g.:
EN 60601-1:2006 + Cor.:2010 + A1:2013,
IEC 60601-1 Edition 3.1:2012,
EN 61010-1:2010 based on IEC 61010-1:2010 + Cor.:2011.
They must be installed outside of the patient area (a radius of 1.5 m
surrounding the patient).
CAUTION
Low voltages are applied to the coupling box for connecting external
interfaces.
➢ Do not touch the pins of the connectors.
NOTE
The externally connected cables must not be subjected to pulling
stress.
CAUTION
Trip/fall hazard
The installation of the supply cable (cable between the coupling box
and the scanner) may result in a trip hazard.
➢ Lay the cable so that there is no risk of tripping.
➢ Attach the supply line so that it remains fixed at all times.
CAUTION
To maintain electrical safety, the laptop must not be operated within
the patient area (1.5 m around the patient).