Dentsply Sirona
Operating Instructions Primescan Connect
6 Maintenance
6.3 Color calibration
67 97 216 D3801
D3801.201.01.03.02 2022-09
69
Replacing the color calibration set
NOTE
Regularly replacing the color calibration set
In order to achieve optimum results, the color calibration set must be
replaced regularly.
➢ Observe the following:
Please note that the color calibration set
● can only be used in conjunction with Connect SW ≥ 5.
● can only be kept for use for a maximum of 2 years. You can find the
expiry date at the bottom of the color calibration set container.
Previous storage may mean that the period for use has been
reduced to less than 2 years.
● can only be used for one year after the container has been opened.
Write the date that the container was opened on the container after
"Opened on ___________" using a waterproof pen and do not use
after one year.
The color calibration set may no longer be used once either of the two
periods has expired.
The software notifies you that the color calibration set needs to be
replaced with a new set before the color calibration expires.
Once the color calibration set has expired the software notifies you that
a color analysis can only be carried out based on old calibration data.
Please contact your dealer for replacements for the color calibration set.