56
COMPRESSOR USER MANUAL
866-DTE-INFO
dentalez.com
PN: 7717-004A
REPLACE INTAKE FILTER
NOTE: All compressor models use the same intake filter.
TOOLS REQUIRED
Torx 25 Screwdriver
1.
On the control panel LCD display, go to the Technical
screen. Press
MAINTENANCE MODE (FIGURE 53)
.
NOTE:
Turning on maintenance mode disables remote
switching and puts the system in standby mode.
2. Remove intake filter cap by removing two screws
(FIGURE 54A)
.
3. Remove intake filter and discard
(FIGURE 54B)
.
4. Install replacement intake filter and replace intake filter
cap.
5. On the Technical screen, press
SERVICE >
(FIGURE 55)
.
6. On the Service screen, press
FILTER PACK
(FIGURE 56)
.
SECTION V - CARE
MAINTENANCE (CONTINUED)
Service
Runtimes
Sensors
Motors
Alerts
High Pressure Mode
Maintenance Mode
07:33:14 PM
Maintenance Mode
LTE
Settings
Technical
Info
Home
Network
FIGURE 55. PRESS SERVICE >
(Continued on the next page.)
FIGURE 54. REMOVE INTAKE FILTER CAP
INTAKE
FILTER
CAP
SCREWS
(2)
A
INTAKE
FILTER
B
FIGURE 53. ENABLE MAINTENANCE MODE
Service
Runtimes
Sensors
Motors
Alerts
High Pressure Mode
Maintenance Mode
07:33:14 PM
Maintenance Mode
LTE
Settings
Technical
Info
Home
Network
Technical
Select item for service
Motor 1
Motor 2
Filter
Pack
Desiccant
Tower
Pressure
Sensor
Ambient
Sensor
Dew Point
Sensor
Service
Desiccant
Cartridge
FIGURE 56. PRESS FILTER PACK