39
MAINTENANCE MODE > SERVICE > DESICCANT CARTRIDGE
TECHNICAL SCREEN (CONTINUED)
SECTION IV - OPERATION
Service
Replace
Desiccant Cartridge
Current Runtime
Max Runtime
Remaining Runtime
WARNING: Replacing item will reset Current Runtime to 0 and update
the serial number (if applicable). This action CANNOT BE REVERSED.
Please confirm item name and serial number above before continuing.
20 hrs
2000 hrs
1980 hrs
Service
Replace
Desiccant Cartridge
Current Runtime
Max Runtime
Remaining Runtime
WARNING: Replacing item will reset Current Runtime to 0 and update
the serial number (if applicable). This action CANNOT BE REVERSED.
Please confirm item name and serial number above before continuing.
20 hrs
2000 hrs
1980 hrs
Replace Desiccant Cartridge?
OK
Cancel
FIGURE 29. DESICCANT CARTRIDGE REPLACEMENT SCREENS
CANCEL
UPDATES
ACCEPT
UPDATES
Service
Replace
Desiccant Cartridge
Current Runtime
Max Runtime
Remaining Runtime
WARNING: Replacing item will reset Current Runtime to 0 and update
the serial number (if applicable). This action CANNOT BE REVERSED.
Please confirm item name and serial number above before continuing.
20 hrs
2000 hrs
1980 hrs
Technical
Select item for service
Motor 1
Motor 2
Filter
Pack
Desiccant
Tower
Pressure
Sensor
Ambient
Sensor
Dew Point
Sensor
Service
Desiccant
Cartridge
SELECT
SERVICE
ITEM
PRESS
REPLACE
RUNTIMES
After replacing the desiccant cartridge, the current and remaining runtimes will need to be reset. On the Service
screen, press
DESICCANT CARTRIDGE
. This will open the Desiccant Cartridge screen showing current, max and
remaining runtimes. Press
REPLACE
. Press
OK
to replace desiccant cartridge and return to Service screen, or press
CANCEL
to cancel the runtimes reset and return to the Desiccant Cartridge screen
(FIGURE 29)
.